The MGISEQ-2000RS FCL sequencing kit features high throughput and low cost per Gb; however, in some situations users don't have enough samples to fill all the lanes, resulting in a much higher run cost. To provide a flexible package, we've developed the MGISEQ-2000RS FCS Sequencing Reagent Kit, which has lower output to make it easier to fully saturate the lanes, while also increasing run speed and flexibility to allow for different combinations of low and medium throughput sequencing. Run more, run faster!
There’s a new addition to the MGISEQ-2000RS series – a faster and more flexible MGISEQ-2000RS high-throughput rapid sequencing reagent kit.Various sample and library types tested by early access users include: tissue, FFPE, cfDNA, gDNA, multiplex PCR, hybridization products, tumor standards of different mutation frequencies, plasma free DNA capture libraries, ctDNA capture libraries. Overall both data output and quality were well above expectations.
Figure 4 Early User FCS PE150 application library data performance
Comments:
Research Science:
The data amount is smaller, there’s more flexibility in arranging the machine runs, and shorter experiment
CWBIO:
We’ve been using PE100 of on the mgiseq-2000 platform, the PE150 of FCS can be especially suitable for some sample mutations that must be detected with a longer read length while also ensuring the same delivery time. The overall quality of the data is quite good, and the amount of data is more flexible, fully meeting the testing needs of our clinical samples.
JingKe Biotech:
The data volume and quality output of the FCS slides are in line with expectations, and are suitable for small and medium-sized laboratories to operate in a more cost effective and autonomous manner.
Gene+:
The data output and quality are in line with expectations, and we are excitedly anticipating the official release.
As a versatile high-throughput gene sequencer, MGISEQ-2000 can be widely used for various applications such as WGS, WES, RNA, WGBS and Panel. However with the increasing variety of users and applications, many companies and laboratories have a strong demand for reagents with smaller data volumes, and higher run flexibility. To this end, MGI has developed a high-throughput rapid sequencing reagent kit for the comprehensive sequencer MGISEQ-2000. The reagent kit has undergone several rounds of internal testing, as well as data quality and sequencing speed optimizations. It has shown excellent performance in key areas such as data accuracy, mutation detection rate, and stability. It was also praised by our early test users. After a long process of trial and error, the company has officially launched a high-throughput rapid sequencing reagent kit.
MGISEQ-2000RS high-throughput rapid sequencing reagent kit uses the FCS slide type, which has a smaller throughput than the FCL slide. At the same time, there are also adjustments to some key components of the sequencing reagent such as reducing the optic capture area, which greatly shortens the sequencing cycle.
In the near future, MGI will also release the FCS exclusive platform MGISEQ-2000RS FAST, a faster, more flexible, and economical dual-slide mid-throughput sequencer, enabling more small and medium-sized enterprises and laboratories to be able to develop their own sequencing capabilities.
(Note: MGISEQ-2000RS high-throughput rapid sequencing reagent kit, hereinafter referred to as FCS reagent kit.)
I.The remarkable features of FCS reagent
Faster Seuqencing Speed
Under the same read length, the FCS slides are 50% faster than the FCL slides. For running PE100, the FCS slide only took about 26 hours. For PE150 it takes only about 37 hours.
No Wasted Lanes
The FCS single-slide is only 550M reads, making it much easier to saturate the all the lanes with less number of samples, eliminating waste and lowering the total operational cost of a single run.
High Data Quality
More speed and the same high quality data. FCS data quality is equivalent to that of FCL.
Cross-platform compatibility
In addition to being compatibile with the existing MGISEQ-2000RS, the FCS exclusive model MGISEQ-2000RS FAST will also be released soon.
MGISEQ-2000RS FAST
Fcs slide
II. Test Performance
We tested the performance of the FCS reagent kit for PE150 sequencing on four MGISEQ-2000RS sequencers. This test had a total of 12 slides or 24 lanes, including 16 standard lanes of the internal standard library, as well as various application types such as tumor detection, WGS, and WES. The reults are as follows:
Figure 1 Standard library sample FCS PE150 data performance
Total reads>300M/lane,CV=2.27%
Q30>88%,CV=0.95%
Split Rate>98%,CV=0.18%
The data showed that the FCS PE150 sequencing reagent kit’s performance on the internal standard library were within expectations in terms of data output, data quality, and resolution.
At the same time, we also tested the performance of the FCS reagent kit for PE100 sequencing.The test was carried out on two MGISEQ-2000RS sequencers, which were divided into five slides, or 10 lanes, including 9 internal standard lanes, and application types including tumor detection. The results are as follows:
Figure 2 Standard library sample FCS PE100 data performance
Q30>88%,CV=0.91%
Split Rate>98%,CV=0.36%
The data showed that the FCS PE100 sequencing reagent kit’s performance on the internal standard library were within expectations in terms of data volume, data quality and resolution.
Table 1 Downstream analysis results of tumor samples
Detection rate |
FCL-A |
FCL-B |
FCL-C |
FCS-A |
FCS-B |
FCS-C |
3%-100% |
0.9784 |
0.989 |
0.989 |
0.996 |
1 |
1 |
1%-3% |
0.9153 |
0.9272 |
0.9272 |
0.887 |
0.923 |
0.923 |
0%-1% |
0.793 |
0.801 |
0.801 |
0.736 |
0.821 |
0.821 |
Number of samples |
3 |
3 |
3 |
5 |
5 |
5 |
The results showed that FCS was slightly better than FCL in the detection of high-frequency variants (>3%), and the detection rate of all samples reached 99%. The detection results for the intermediate frequency variants (1%-3%) were within expectations.
IV. Comprehensive evaluation
After several rounds of testing, the MGISEQ-2000RS high-throughput rapid sequencing reagent kit (FCS reagent kit) performed well in areas such as data accuracy, mutation detection rate, and stability. The FCS kit is a great addition for any users looking to increase their run flexibility, reduce run costs, and maintain the same level of high quality data as the FCL kits.
The MGISEQ-2000RS high-throughput rapid sequencing reagent kit is available now, in FCS PE100 and FCS PE150 configurations. Order yours now and let’s run more, and run faster!
Product Name (English) |
version number |
Item number |
MGISEQ-2000RS High-throughput Rapid Sequencing Set (FCS PE100) |
V1.0 |
1000013155 |
MGISEQ-2000RS High-throughput Rapid Sequencing Set (FCS PE150) |
V1.0 |
1000011718 |