MGISP-NE384
MGISP-NE384 is a high-throughput automated nucleic acid extractor adopting magnetic rod technology. Together with nucleic acid extraction kit, MGISP-NE384 is able to extract and purify nucleic acid from 96/192/288/384 samples. HEPA filtration system and UV lamp, ensure a safe working environment for nucleic acid extraction and lab technicians.
Features
Flexible: Mixing and magnetic module Z axis independent control, 1~4 modules running choice for 96/192/288/384 samples.
|
Efficient: Moving and processing magnetic particles, extracting and purifying nucleic acid from 384 samples.
|
Safe: HEPA filtration system and UV lamp, avoiding cross-contamination and ensuring a safe working environment.
|
Stable: Reliable hardware design,96-well plate reagent and disposable tip, minimizing operational errors.
|
Performance Parameter
MGISP-NE384 | ||
---|---|---|
Capacity | 384 samples | |
Processing Volume |
20 µl~1000 µl |
|
Working Position | 4x6 96-well plate working position | |
Temperature Control, Lysis/Elution Well |
From +5℃ above ambient temperature to 115℃ |
|
Temperature Precision |
≤±1℃ |
|
Mixing Module |
Z axis independent control, 1~4 modules running choice |
|
UV Lamp |
More than 100,000 μW.s/cm2 | |
HEPA Filtration System | 99.99% at 3 µm | |
Operating Interface | 23-inch built-in touch screen, Windows software | |
Program Management | Create, edit, and delete program |
System Parameter
- Working Environment
Temperature: 19 ℃ ~ 25 ℃
Relative humidity: 20% RH ~ 80% RH
Atmospheric pressure: 80 kPa ~ 106 kPa
- Power Supply & Dimensions & Weight
Power Supply: 200 V~240 V, 50/60 Hz, 2350 W
Dimensions: length:1220 mm; width: 742 mm; height: 960 mm